Experimental Substantiation of a Number of Requirements for Hardware and Methodology of Non-Contact Photoplethysmography Based on Video Image Analysis

Authors: Gerzhik A.A., Raznitsyna I.A. Published: 27.12.2021
Published in issue: #4(137)/2021  
DOI: 10.18698/0236-3933-2021-4-122-138

 
Category: Instrument Engineering, Metrology, Information-Measuring Instruments and Systems | Chapter: Optical and Optoelectronic Instruments and Complexes  
Keywords: digital camera, non-contact photoplethysmography, hemodynamics, in vivo, RGB

Experimental assessment of the possibility of using a scientific video camera for realization of non-contact photoplethysmography is carried out. In view of the wide spread of digital cameras in endoscopic units photoplethysmography based on video-image analysis is an inexpensive and promising method for solving problems of medical diagnostics. A number of requirements to the camera parameters ensuring the specified level of the registered signals coded by RGB values, to the external illumination and video image postprocessing algorithms was substantiated. It was found that at signal levels on the green channel of not less than 130 and not more than 220 a. u. (for a camera with 8-bit color coding depth), the highest level of useful pulse signal is provided. It was shown that any source of light of equivalent color temperature in the range of 3500--6500 K as well as green LEDs can be used as a light source to obtain a high-quality signal of non-contact photoplethysmography. It is demonstrated that the signal has higher signal/noise ratio when the averaging area is larger than 40 × 40 pixels than when averaging over a smaller group of pixels. The obtained results can be used for the implementation of non-contact photoplethysmography to study local blood flow based on video image analysis This work was supported by the Foundation for Assistance to Small InnovativeEnterprises in Science and Technology (contract no. 13938GU/2019)

References

[1] Challoner A.V.J. Photoelectric plethysmography for estimating cutaneous blood flow. In: Rolfe P. (ed.). Non-invasive Physiological Measurements. Vol. 1. Academic, 1979, pp. 125--151.

[2] Rogatkin D.A. Physical foundations of contemporary optical methods for the study of microhemodynamics in vivo. Lecture. Meditsinskaya fizika [Medical Physics], 2017, no. 4, pp. 75--93 (in Russ.).

[3] Semchuk I.P., Muravskaya N.P., Samorodov A.V. [Experimental study on optical methods of contactless photoplethysmography]. Fizika i radioelektronika v meditsine i ekologii [Physics and Radioelectronics in Medicine and Ecology]. Moscow, Grafika Publ., 2018, pp. 86--89 (in Russ.).

[4] Zil’ber A.P., ed. Etyudy meditsinskogo prava i etiki [Studies on medical law and ethics]. Moscow, MEDpress-inform Publ., 2008.

[5] Tuchin V.V., ed. Opticheskaya meditsinskaya diagnostika. T. 1 [Optical medical diagnostics. Vol. 1]. Moscow, FIZMATLIT Publ., 2007.

[6] Wang D., Yang X., Liu X., et al. Detail-preserving pulse wave extraction from facial videos using consume-level camera. Biomed. Opt. Express, 2020, vol. 11, no. 4, pp. 1876--1891. DOI: https://doi.org/10.1364/boe.380646

[7] Nishidate I., Maeda T., Niizeki K., et al. Estimation of melanin and hemoglobin using spectral reflectance images reconstructed from a digital RGB image by the Wiener estimation method. Sensors, 2013, vol. 13, no. 6, pp. 7902--7915. DOI: https://doi.org/10.3390/s130607902

[8] Jones G., Clancy N.T., Arridge S., et al. Inference of tissue haemoglobin concentration from stereo RGB. Proc. MIAR, 2016, pp. 50--58. DOI: https://doi.org/10.1007/978-3-319-43775-0_5

[9] Wang F. Linearity research of a CMOS image sensor. Delft, TU Electronic Instrumentation, 2018. DOI: https://doi.org/10.4233/uuid:9d79cf6d-19a5-4f0f-a01e-6573f8e1b2ce

[10] De Haan G., Jeanne V. Robust pulse rate from chrominance-based rPPG. IEEE Trans. Biomed. Eng., 2013, vol. 60, no. 10, pp. 2878--2886. DOI: https://doi.org/10.1109/TBME.2013.2266196

[11] Janesick J.R. Scientific charge-coupled devices. Bellingham, SPIE Press, 2001.

[12] Hytti H.T. Characterization of digital image noise properties based on RAW data. Proc. SPIE, 2006, vol. 6059, art. 60590A. DOI: https://doi.org/10.1117/12.640500

[13] NOIP1SN1300A. PYTHON 1.3/0.5/0.3 MegaPixels global shutter CMOS image sensors. Denver, Phoenix, Literature Distribution Center for ON Semiconductor, 2016.

[14] Chen D.Y., Wang J.J., Lin K.Y., et al. Image sensor-based heart rate evaluation from face reflectance using Hilbert --- Huang transform. IEEE Sensors J., 2014, vol. 15, no. 1, pp. 618--627. DOI: https://doi.org/10.1109/JSEN.2014.2347397

[15] Wang C., Pun T., Chanel G. A comparative survey of methods for remote heart rate detection from frontal face videos. Front. Bioeng. Biotechnol., 2018, vol. 6, art. 33. DOI: https://doi.org/10.3389/fbioe.2018.00033

[16] Kamshilin A.A., Teplov V., Nippolainen E., et al. Variability of microcirculation detected by blood pulsation imaging. PloS One, 2013, vol. 8, no. 2, art. e57117. DOI: https://doi.org/10.1371/journal.pone.0057117

[17] Verkruysse W., Svaasand L.O., Nelson J.S. Remote plethysmographic imaging using ambient light. Opt. Express, 2008, vol. 16, no. 26, pp. 21434--21445. DOI: https://doi.org/10.1364/OE.16.021434

[18] Lapitan D.G., Raznitsyn O.A. A method and a device prototype for noninvasive measurements of blood perfusion in a tissue. Instrum. Exp. Tech., 2018, vol. 61, no. 5, pp. 745--750. DOI: https://doi.org/10.1134/S0020441218050093